清洗步驟:
Cleaning steps:
1、預(yù)處理:內(nèi)鏡使用后即刻使用濕紗布擦拭外表污漬,置于容器中送清洗消毒室,同時(shí)檢查鏡子等有無(wú)測(cè)漏、損壞;
1. Pre treatment: Immediately after using the endoscope, use a damp gauze to wipe off external stains, place it in a container and send it to a cleaning and disinfection room. At the same time, check the mirror for leaks or damage;
2、水洗:用流動(dòng)水徹底清洗內(nèi)鏡的各個(gè)部件,以高壓水槍徹底清洗各管腔;
2. Water washing: Thoroughly clean all components of the endoscope with flowing water, and thoroughly clean each lumen with a high-pressure water gun;
3、酶洗:內(nèi)鏡清洗擦干后,置于酶洗槽中浸泡。
3. Enzyme washing: After cleaning and drying the endoscope, soak it in an enzyme washing tank.
4、清洗:可拆卸部分必須拆開清洗,并用超聲清洗器清洗5-10分鐘;器械的軸節(jié)部、彎曲部、官腔內(nèi)用軟毛刷刷洗。用毛刷刷洗時(shí)注意避免劃傷鏡面。
4. Cleaning: The detachable part must be disassembled for cleaning and cleaned with an ultrasonic cleaner for 5-10 minutes; Use a soft bristle brush to brush and wash the axis, curvature, and internal cavity of the instrument. When brushing with a brush, be careful not to scratch the mirror surface.
5、消毒與滅菌:按照內(nèi)鏡的規(guī)格要求選擇正確的消毒方式進(jìn)行消毒;(負(fù)壓吸引器、超聲吸引器、高壓水槍、干燥設(shè)備、計(jì)時(shí)器等);
5. Disinfection and sterilization: Select the correct disinfection method according to the specifications of the endoscope for disinfection; (Negative pressure suction device, ultrasonic suction device, high-pressure water gun, drying equipment, timer, etc.);
拆裝過(guò)的鏡子建議采用低溫等離子進(jìn)行消毒;
It is recommended to use low-temperature plasma for disinfection of disassembled mirrors;
根據(jù)內(nèi)鏡在人體內(nèi)的使用部位不同進(jìn)行消毒與滅菌,比如,進(jìn)入人體自然通道的官腔粘膜接觸的內(nèi)鏡及其附件,應(yīng)用前要采取過(guò)高水平的消毒滅菌。
Disinfection and sterilization should be carried out according to the different parts of the endoscope used in the human body. For example, endoscopes and their accessories that come into contact with the natural channels of the human body should be disinfected and sterilized at a high level before application.
The endoscope equipment maintenance personnel will now talk about several common troubleshooting repairs:
鉗子管漏水:多數(shù)是由于活檢鉗等附件,使用不當(dāng)造成活檢管道破損,在視野不模糊時(shí),C級(jí)修理
Water leakage in the forceps tube: Most cases are caused by improper use of biopsy forceps and other accessories, resulting in damage to the biopsy tube. When the visual field is not blurred, C-level repair is required
導(dǎo)管束斷:使用例數(shù)多,彎曲管不良造成,C+導(dǎo)光束級(jí)修理
Broken catheter bundle: caused by numerous use cases and poor bending of the tube, C+beam level repair
鋼絲前端斷:使用粒數(shù)多造成,C級(jí)修理
Broken front end of steel wire: caused by excessive use of particles, C-level repair
彎曲角度不足:鋼絲使用一定時(shí)間后鋼絲變長(zhǎng),D級(jí)修理,拆開后剪短相應(yīng)方向的鋼絲即可,但需判別彎曲管是否良好
Insufficient bending angle: After using the steel wire for a certain period of time, the steel wire becomes longer. For D-level repair, simply cut the steel wire in the corresponding direction after dismantling, but it is necessary to determine whether the bent pipe is in good condition
彎曲橡皮漏水:碰傷造成比較多,D級(jí)修理更換專用橡皮即可
Leakage of bent rubber: more damage caused by collisions, replacement with specialized rubber for D-level repair is sufficient
噴嘴堵塞:清洗不干凈,C級(jí)修理更換噴嘴
Nozzle blockage: cleaning not clean, C-level repair and nozzle replacement
導(dǎo)像束黑點(diǎn)多或龜裂:纖維鏡使用粒數(shù)多,A級(jí)修理
Multiple black spots or cracks in the guiding image beam: Fiber mirrors with a large number of particles are used for A-level repair
具體故障需經(jīng)漏水測(cè)試,烘干,拆鏡檢查仔細(xì)判別才能定如何修理
Specific faults need to be carefully identified through water leakage testing, drying, and mirror inspection before determining how to repair them
內(nèi)窺鏡的清洗消毒操作及常見故障修理相關(guān)事項(xiàng)就講解到這里了,希望能夠給您幫助,更多事項(xiàng)就來(lái)我們網(wǎng)站
http://m.tukouzhao.cn進(jìn)行咨詢!
That's all for the cleaning and disinfection operations of endoscopes, as well as the repair of common faults. We hope to provide you with assistance. For more information, please visit our website http://m.tukouzhao.cn Consulting!