內(nèi)窺鏡的相關(guān)參數(shù)及故障分析

來源:http://m.tukouzhao.cn/ 發(fā)布時間:2023-04-06 瀏覽量:0

隨著科學(xué)技術(shù)的發(fā)展,醫(yī)用內(nèi)窺鏡已經(jīng)被廣泛的應(yīng)用于醫(yī)療領(lǐng)域,它是人類窺視、治療人體內(nèi)器官的重要工具之一。下面內(nèi)窺鏡維修公司來講講內(nèi)窺鏡的相關(guān)參數(shù)及故障:
With the development of science and technology, medical endoscopes have been widely used in the medical field. They are one of the important tools for humans to observe and treat internal organs of the human body. Below, the endoscope maintenance company will talk about the relevant parameters and faults of the endoscope:
(1)光源系統(tǒng)常見的故障
(1) Common faults in light source systems
(a)光線發(fā)暗。排除了是光柵遮擋的原因外,基本可以肯定是燈泡接近燈泡使用壽命,更換新燈泡即可。氙光燈泡的使用壽命一般設(shè)計在500小時,如果光源系統(tǒng)有壽命計時裝置,這個問題就很容易解決。
(a) The light is dim. Excluding the cause of grating obstruction, it is generally certain that the bulb is approaching its lifespan and can be replaced with a new one. The service life of xenon light bulbs is generally designed to be 500 hours. If the light source system has a lifespan timer, this problem can be easily solved.
(b)開機燈泡不亮。如果排除了燒掉燈泡這個原因的話,那么問題就常常出現(xiàn)在主機高壓點火電路部分,因為開機的時候機器會產(chǎn)生上萬伏特的脈沖高壓,元件會老化的特變快。注意此時不能用萬用表去檢查脈沖高壓。
(b) The light bulb does not turn on when turned on. If the cause of burning out the light bulb is ruled out, the problem often occurs in the high-voltage ignition circuit of the host, because when the machine is turned on, it will generate tens of thousands of volts of pulse high voltage, and the components will age very quickly. Note that a multimeter cannot be used to check the pulse high voltage at this time.
(2)人工氣腹系統(tǒng)常見的故障
(2) Common faults of artificial pneumoperitoneum system
(a)壓力顯示不正確。這多是氣體壓力傳感器損壞,如果機器設(shè)計有一組多個壓差壓力傳感器同時工作的,一般一個老化或者污染無法工作的時候,我們要把這組傳感器全部更換,否則只換其中一個,其他的同組的傳感器壽命也接近臨界了。
(a) The pressure display is incorrect. This is mostly due to damage to the gas pressure sensor. If the machine is designed with multiple differential pressure sensors working simultaneously, and one sensor is generally aged or contaminated and cannot work, we need to replace all of these sensors. Otherwise, only one of them will be replaced, and the lifespan of other sensors in the same group is also approaching the critical point.
(b)設(shè)備漏氣。很多情況是器械密封圈老化而引起的。但是也有設(shè)備機械減壓閥漏氣和電磁閥損壞的原因,這個比較容易判斷出并對其進行修復(fù)。
(b) Equipment leaks air. Many cases are caused by the aging of the instrument sealing ring. However, there are also reasons for equipment mechanical pressure reducing valve leakage and electromagnetic valve damage, which are relatively easy to identify and repair.
內(nèi)窺鏡維修公司
(3)電凝電切系統(tǒng)常見的故障
(3) Common faults in electrocoagulation and electrocautery systems
(a)設(shè)備配件損壞。如腳踏開關(guān),負極板導(dǎo)線,連接導(dǎo)線等常用的易損配件。
(a) Equipment accessories are damaged. Commonly used vulnerable accessories such as foot switches, negative electrode wires, connecting wires, etc.
(b)電切、電凝、雙極故障。在設(shè)備內(nèi)部,這些部分都由不同的電路模塊組成。我們可以根據(jù)其表現(xiàn)的故障現(xiàn)象對相應(yīng)的模塊電路進行維修或更換。
(b) Electric cutting, coagulation, bipolar faults. Inside the device, these parts are composed of different circuit modules. We can repair or replace the corresponding module circuits based on the fault symptoms they exhibit.
(4)沖洗吸引系統(tǒng)常見故障
(4) Common faults of flushing suction system
(a)壓力或者吸力不足。設(shè)備中的雙向泵老化,或者負壓瓶漏氣。
(a) Insufficient pressure or suction. The bidirectional pump in the equipment is aging, or the negative pressure bottle is leaking.
(b)設(shè)備無法開機,一般都為液體進入設(shè)備,在沒燒開相關(guān)部件的情況下清理干凈被錯吸的液體就可以了。
(b) The device cannot be turned on, usually due to liquid entering the device. It is sufficient to clean the mistakenly aspirated liquid without burning the relevant components.
為了盡可能減少設(shè)備故障的發(fā)生,應(yīng)設(shè)立專用的手術(shù)間,以減少儀器的搬動與損壞,并指派??谱o士使用和保養(yǎng),嚴格執(zhí)行操作規(guī)程,抓好手術(shù)前檢查、術(shù)中故障排除、術(shù)后及時維護保養(yǎng)3個環(huán)節(jié)。如果是設(shè)備出現(xiàn)問題,首先是采取應(yīng)急維修,這樣能解決占故障85%左右的中小故障。更多相關(guān)內(nèi)容就來我們網(wǎng)站http://m.tukouzhao.cn咨詢了解!
In order to minimize the occurrence of equipment failures as much as possible, a dedicated operating room should be established to reduce the handling and damage of instruments, and specialized nurses should be assigned to use and maintain them. Operating procedures should be strictly followed, and three aspects should be emphasized: preoperative inspection, intraoperative troubleshooting, and timely postoperative maintenance. If there is a problem with the equipment, the first step is to take emergency maintenance, which can solve about 85% of small and medium-sized faults. For more related content, come to our website http://m.tukouzhao.cn Consultation and understanding!
s
新聞資訊
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設(shè)備有限公司
  • 匠仁醫(yī)療山東濟南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話:13153199508
  • 地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室
  • 匠仁醫(yī)療湖南長沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區(qū)勞動?xùn)|路820號恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室  湖南省長沙市雨花區(qū)勞動?xùn)|路820號恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號