歡迎來到匠仁醫(yī)療設(shè)備有限公司網(wǎng)站!

更多 > 新聞資訊
公司新聞
行業(yè)知識
檢測答疑
圖標 首頁 >> 新聞資訊 > 行業(yè)知識

行業(yè)知識

內(nèi)窺鏡設(shè)備的4種系統(tǒng)常見故障

來源:http://m.tukouzhao.cn/ 發(fā)布時間:2022-11-12 瀏覽量:0

醫(yī)用內(nèi)窺鏡是一種常用的醫(yī)療器械,由可彎曲部分、光源及一組鏡頭組成。經(jīng)人體的天然孔道,或是經(jīng)過手術(shù)做的小切口進入人體體內(nèi)。那么您知道內(nèi)窺鏡設(shè)備的相關(guān)參數(shù)及故障包含什么嗎?下面內(nèi)窺鏡設(shè)備維修人員給您講解。
Medical endoscope is a commonly used medical instrument, which consists of a flexible part, a light source and a group of lenses. It enters the human body through natural channels or small incisions made by surgery. Do you know what the relevant parameters and faults of the endoscope equipment include? The endoscope equipment maintenance personnel will explain to you below.
(1)光源系統(tǒng)常見的故障
(1) Common faults of light source system
光線發(fā)暗。排除了是光柵遮擋的原因外,基本可以肯定是燈泡接近使用壽命,更換新燈泡則可。氙光燈泡的使用壽命一般設(shè)計在500小時,如果光源系統(tǒng)有壽命計時裝置,這個問題就很容易解決。
The light is dim. It is ruled out that the light bulb is close to its service life, and it can be replaced with a new light bulb. The service life of xenon light bulb is generally designed to be 500 hours. If the light source system has a life timer, this problem can be easily solved.
開機燈泡不亮。如果排除了燒掉燈泡這個原因的話,那么問題常常就出在主機高壓點火電路部分,因為開機的時候機器會產(chǎn)生上萬伏特的脈沖高壓,元件會老化得特別快。注意此時不能用萬用表去檢查脈沖高壓。
The power on bulb does not light up. If the reason of burning out the bulb is ruled out, the problem is often in the high-voltage ignition circuit of the host, because the machine will generate tens of thousands of volts of pulse high voltage when it is turned on, and the components will age very quickly. Note that you cannot use a multimeter to check the pulse high voltage at this time.
(2)人工氣腹系統(tǒng)常見的故障
(2) Common faults of artificial pneumoperitoneum system
壓力顯示不正確。這多是氣體壓力傳感器損壞,如果機器設(shè)計有一組多個壓差壓力傳感器同時工作的,一般一個老化或者污染無法工作的時候,我們要把這組傳感器全部更換,否則只換其中一個,其他的同組的傳感器壽命也接近臨界了。
The pressure display is incorrect. This is mostly due to the damage of gas pressure sensors. If the machine is designed to have a group of multiple differential pressure sensors working at the same time, generally when one of them is aging or polluted and cannot work, we should replace all of these sensors, otherwise only one of them will be replaced, and the life of other sensors of the same group will also be near the critical point.
流量不準確。氣體流量傳感器損壞更換即可。有時候是通過壓差來計算流量的,這時候我們就要更換新的壓差傳感器。
The flow is not accurate. If the gas flow sensor is damaged, replace it. Sometimes the flow is calculated by differential pressure. At this time, we need to replace a new differential pressure sensor.
設(shè)備漏氣。很多情況是器械密封圈老化而引起的。但也有設(shè)備機械減壓閥漏氣和電磁閥損壞的原因,這個比較容易判斷出并對其進行修復(fù)。
The equipment leaks air. Many cases are caused by the aging of the device sealing ring. However, there are also reasons for leakage of mechanical pressure reducing valve and damage of solenoid valve, which can be easily identified and repaired.
內(nèi)窺鏡設(shè)備維修
(3)液體加壓系統(tǒng)常見的故障
(3) Common faults of liquid pressurizing system
蠕動泵損壞。
Peristaltic pump is damaged.
液體壓力傳感器損壞。其原理基本與氣體壓力傳感器相同。
The liquid pressure sensor is damaged. Its principle is basically the same as that of the gas pressure sensor.
設(shè)備內(nèi)部電路損壞。
The internal circuit of the equipment is damaged.
(4)電凝電切系統(tǒng)常見的故障
(4) Common faults of electric coagulation and electric cutting system
設(shè)備配件損壞。如腳踏開關(guān),負極板導線,連接導線等常用的易損配件。
The equipment accessories are damaged. Such as foot switch, negative plate wire, connecting wire and other commonly used vulnerable accessories.
電切、電凝,雙極故障。在設(shè)備內(nèi)部,這些部分都由不同的電路模塊組成。我們可以根據(jù)其表現(xiàn)的故障現(xiàn)象對相應(yīng)的模塊電路進行維修或更換。
Electric cutting, electrocoagulation, bipolar fault. Inside the equipment, these parts are composed of different circuit modules. We can repair or replace the corresponding module circuit according to the fault phenomenon.
如果是設(shè)備出問題,應(yīng)該先采取應(yīng)急維修,因為這樣能解決占故障85%左右的中小故障。而在實際維修的時候難解決的問題是更換配件的訂購。更多相關(guān)內(nèi)容就來我們網(wǎng)站http://m.tukouzhao.cn進行咨詢了解吧!
If there is a problem with the equipment, emergency maintenance should be carried out first, because this can solve the small and medium faults that account for about 85% of the faults. The problem that is difficult to solve in actual maintenance is the ordering of replacement parts. More relevant content will come to our website http://m.tukouzhao.cn Consult and understand!
s
新聞資訊
聯(lián)系我們
  • 公司公司:匠仁醫(yī)療設(shè)備有限公司
  • 匠仁醫(yī)療山東濟南維修部
  • 聯(lián)系人:樊經(jīng)理
  • 電話:13153199508
  • 地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室
  • 匠仁醫(yī)療湖南長沙維修部
  • 聯(lián)系人:李經(jīng)理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區(qū)勞動東路820號恒大綠洲14棟2409室
新聞資訊

樊經(jīng)理:13153199508     李經(jīng)理:13873135765

公司地址:山東省濟南市槐蔭區(qū)美里東路3000號德邁國際信息產(chǎn)業(yè)園6號樓101-2室  湖南省長沙市雨花區(qū)勞動東路820號恒大綠洲14棟2409室

     

魯公網(wǎng)安備 37010402441281號